Summary
The dataset -- including 4D fan and cone beam CT technique and collection protocol, delineation protocol, an -- is most fully described in detail in Balik et al.1 Briefly, 4D-FBCT images were acquired on a 16-slice helical CT scanner (Brilliance Big Bore, Philips Medical Systems, Andover, MA) as respiration-correlated CTs with 10 breathing phases (0 to 90%, phase-based binning) and 3 mm slice thickness. 4D-FBCT images were acquired during simulation, prior to therapy, and used for therapy planning. In 14 of the 20 subjects, 4D-FBCTs were also acquired on the same scanner weekly during therapy. 4D-CBCT images were acquired on a commercial CBCT scanner (On-Board Imager™, Varian Medical Systems, Inc.). An external surrogate (Real-time Position Management, Varian Medical Systems, Inc.) was integrated into the CBCT acquisition system to stamp each CBCT projection with the surrogate respiratory signal through in-house software and hardware tools. Approximately 2500 projections were acquired over a period of 8-10 minutes in half-fan mode with half bow-tie filter. The technique was 125 kVp, 20 mA, and 20 ms in a single 360° slow gantry arc. Using the external surrogate, the CBCT projections were sorted into 10 breathing phases (0 to 90%, phase-based binning) and reconstructed with an in-house FDK reconstruction algorithm. |