Summary
This collection consists of 17 CT scans of the Credence Cartridge Radiomics (CCR) phantom, which was designed for use in studies of texture feature robustness. The scans were acquired at four medical centers using each center’s chest protocol and were taken using GE (7 scans), Philips (5 scans), Siemens (2 scans), and Toshiba (3 scans) scanners. The CCR phantom has 10 cartridges, each with a unique texture, Fig 1. The first four cartridges are 3D printed ABS plastic with 20%, 30%, 40%, and 50% honeycomb fill, and they provide regular, periodic textures. The next three cartridges provide natural textures: sycamore wood, cork, and extra dense cork. A cartridges of shredded rubber particles provides textures similar to those of non-small cell lung cancer. The ninth cartridge is solid, homogenous acrylic and provides a minimal texture control. Finally, the 10th cartridge is 3D printed plaster has the highest electron density (400 – 600 HU) and is intended to more similar to bone.
In addition to the DICOM images for the 17 scans, this collection also contains two sets of contours as DICOM RT structure files. The first set provides 8x8x2 cm3 contours for each cartridge in each scan. The second set provides 16 adjacent 2x2x2 cm3 contours for each cartridge in each scan.
Excerpt |
---|
Researchers studying radiomics will be able to evaluate features for robustness across a variety of scanners. Features can be calculated using the researchers own software or third party software such as IBEX (imaging biomarker explorer).
Related publications: http://journals.lww.com/investigativeradiology/Abstract/2015/11000/Measuring_Computed_Tomography_Scanner_Variability.3.aspx
The following paper was generated on different imaging modalities but the same phantom, this is a related but independent paper with a different set of authors: http://tinyurl.com/zm7tr7p
Acknowledgments
This data set was provided to TCIA by Authors: Mackin, Dennis; Fave, Xenia; Zhang, Lifei; Fried, David; Yang, Jinzhong; Taylor, Brian; Rodriguez-Rivera, Edgardo; Dodge, Cristina; Jones, Aaron Kyle; and Court, Laurence.
Localtab Group |
---|
Localtab |
---|
active | true |
---|
title | Data Access |
---|
| Data AccessChoosing the Download option will provide you with a file to launch the TCIA Download Manager to download the entire collection. If you want to browse or filter the data to select only specific scans/studies please use the Search By Collection optionClick the Download button to save a ".tcia" manifest file to your computer, which you must open with the NBIA Data Retriever . Click the Search button to open our Data Portal, where you can browse the data collection and/or download a subset of its contents. Data Type | Download all or Query/Filter |
---|
Images Images and Radiation Therapy Structures (DICOM, 1.33 GB) | |
Click the Versions tab for more info about data releases. |
Localtab |
---|
title | Detailed Description |
---|
| Detailed Description | |
---|
Modalities | CT, RTSRUCT | Number of | PatientsParticipants | 17 | Number of Studies | 17 | Number of Series | 51 | Number of Images | 2672 | Image Size (GB) | 1.33 |
Acquisition parameters for the phantom scans in this Collection: Scan | Manufacturer | Model | Kernel | Type | Slice Thickness (mm) | Pixel (mm) | Spiral Pitch Factor | kVp | Effective mAs | CTDIvol (mGy) | CCR1-GE1 | GE | Discovery CT750 HD | standard | helical | 2.5 | 0.49 | 0.98 | 120 | 81 | 6.19 | CCR1-GE2 | GE | Discovery CT750 HD | standard | axial | 2.5 | 0.70 | 1.00 | 120 | 300 | |
| CCR1-GE3 | GE | Discovery CT750 HD | standard | helical | 2.5 | 0.78 | 0.98 | 120 | 122 | 9.3 | CCR1-GE4 | GE | Discovery ST | standard | helical | 2.5 | 0.98 | 1.35 | 120 | 143 | 16.3 | CCR1-GE5 | GE | LightSpeed RT | standard | helical | 2.5 | 0.98 | 0.75 | 120 | 1102 | 53.6 | CCR1-GE6 | GE | LightSpeed RT16 | standard | helical | 2.5 | 0.98 | 0.94 | 120 | 367 | 18.8 | CCR1-GE7 | GE | LightSpeed VCT | standard | helical | 2.5 | 0.74 | 0.98 | 120 | 82 | |
| CCR1-P1 | Philips | Brilliance Big Bore | B | helical | 3.0 | 0.98 | 0.94 | 120 | 320 | 17.8 | CCR1-P2 | Philips | Brilliance Big Bore | C | helical | 3.0 | 0.98 | 0.94 | 120 | 369 | 15.8 | CCR1-P3 | Philips | Brilliance Big Bore | B | helical | 3.0 | 1.04 | 0.81 | 120 | 320 | 19.9 | CCR1-P4 | Philips | Brilliance Big Bore | B | helical | 3.0 | 1.04 | 0.81 | 120 | 369 | 19.9 | CCR1-P5 | Philips | Brilliance 64 | B | helical | 3.0 | 0.98 | 0.67 | 120 | 372 | 16.4 | CCR1-S1 | Siemens | Sensation Open | B31s | axial | 3.0 | 0.54 | 1.00 | 120 | 26 - 70 | 1.5 | CCR1-S2 | Siemens | SOMATOM Definition Flash | ['I70f', '2'] | helical | 2.0 | 0.52 | 0.60 | 120 | 17 - 28 | |
| CCR1-T1 | Toshiba | Aquilion | FC18 | helical | 3.0 | 0.63 | 1.11 | 120 | 135 | 4.0 | CCR1-T2 | Toshiba | Aquilion | FC18 | helical | 3.0 | 0.63 | 1.11 | 120 | 135 | 3.8 | CCR1-T3 | Toshiba | Aquilion ONE | FC18 | helical | 3.0 | 0.98 | 0.99 | 120 | 151 | 13.5 |
|
Localtab |
---|
title | Citations & Data Usage Policy |
---|
| Citations & Data Usage Policy This collection is freely available to browse, download, and use for commercial, scientific and educational purposes as outlined in the Creative Commons Attribution 3.0 Unported License. See TCIA's Data Usage Policies and Restrictions for additional details. Questions may be directed to help@cancerimagingarchive.net. Please be sure to include the following citations in your work if you use this data set: Public collection license |
---|
Info |
---|
| Mackin, Dennis; Fave, Xenia; Zhang, Lifei; Fried, David; Yang, Jinzhong; Taylor, Brian; Rodriguez-Rivera, Edgardo; Dodge, Cristina; Jones, Aaron Kyle; and Court, Laurence. (2017). Data From Credence Cartridge Radiomics Phantom CT Scans. The Cancer Imaging Archive. http://doi.org/10.7937/K9/TCIA.2017.zuzrml5b |
Info |
---|
| Clark K, Vendt B, Smith K, Freymann J, Kirby J, Koppel P, Moore S, Phillips S, Maffitt D, Pringle M, Tarbox L, Prior F. The Cancer Imaging Archive (TCIA): Maintaining and Operating a Public Information Repository, Journal of Digital Imaging, Volume 26, Number 6, December, 2013, pp 1045-1057. (paper) |
Other Publications Using This Data TCIA maintains a list of publications which leverage our data. If you have a publication you'd like to add please contact the TCIA Helpdesk . |
Localtab |
---|
| Version 1 (Current): Updated 2017/07/28 Data Type | Download all or Query/Filter |
---|
Images (1.33 GB) | |
|
|